The Mira hormone monitor device firmware accepts a 0x01 write from any BLE central without authentication, causing the device to reboot into bootloader mode. An attacker could cause a denial-of-service condition or disrupt ovulation tracking and fertility monitoring workflow.
https://www.hipaajournal.com/vulnerabilities-mira-hormone-monitor-pulsetto-vagus-nerve-stimulator/
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01
https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json
Published: 2026-08-11
Updated: 2026-08-12
Base Score: 6.1
Vector: CVSS2#AV:A/AC:L/Au:N/C:N/I:N/A:C
Severity: Medium
Base Score: 6.5
Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Severity: Medium
Base Score: 7.1
Vector: CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
Severity: High
EPSS: 0.00197