In the default configuration, the archive accepts DICOM associations from an unregistered calling AE. A remote peer can use C-FIND and C-GET to enumerate and retrieve patient information and medical images. It can also use C-STORE to file fabricated study content under an existing patient identity.
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-288-02
https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-288-02.json
Published: 2026-10-16
Updated: 2026-10-16
Base Score: 9.4
Vector: CVSS2#AV:N/AC:L/Au:N/C:C/I:C/A:N
Severity: High
Base Score: 9.1
Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
Severity: Critical
Base Score: 9.3
Vector: CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N