CVE-2024-48967

critical

Description

The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.

References

https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01

Details

Source: Mitre, NVD

Published: 2024-11-14

Updated: 2026-04-15

Risk Information

CVSS v2

Base Score: 10

Vector: CVSS2#AV:N/AC:L/Au:N/C:C/I:C/A:C

Severity: Critical

CVSS v3

Base Score: 10

Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Severity: Critical

EPSS

EPSS: 0.0013