Description

An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices.

References

https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02

https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html

Details

Source: Mitre, NVD

Published: 2022-04-14

Updated: 2026-06-17

Risk Information

CVSS v2

Base Score: 7.5

Vector: CVSS2#AV:N/AC:L/Au:S/C:N/I:C/A:P

Severity: High

CVSS v3

Base Score: 7.1

Vector: CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:L

Severity: High

EPSS

EPSS: 0.00026