CVE-2020-12040

critical

Description

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.

References

https://www.us-cert.gov/ics/advisories/icsma-20-170-04

Details

Source: Mitre, NVD

Published: 2020-06-29

Updated: 2026-06-17

Risk Information

CVSS v2

Base Score: 5

Vector: CVSS2#AV:N/AC:L/Au:N/C:P/I:N/A:N

Severity: Medium

CVSS v3

Base Score: 9.8

Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Severity: Critical

EPSS

EPSS: 0.00177