CVE-2019-6540

medium

Description

The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.

References

https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01

http://www.securityfocus.com/bid/107544

Details

Source: Mitre, NVD

Published: 2019-03-26

Updated: 2025-05-22

Risk Information

CVSS v2

Base Score: 3.3

Vector: CVSS2#AV:A/AC:L/Au:N/C:P/I:N/A:N

Severity: Low

CVSS v3

Base Score: 6.5

Vector: CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Severity: Medium

EPSS

EPSS: 0.00018