CVE-2019-25724

high

Description

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

References

https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-x-network-based-denial-of-service

https://static.draeger.com/security/download/2019-277-02-M300-VG2x-Security-Advisory.pdf

Details

Source: Mitre, NVD

Published: 2026-06-02

Updated: 2026-06-04

Risk Information

CVSS v2

Base Score: 6.1

Vector: CVSS2#AV:A/AC:L/Au:N/C:N/I:N/A:C

Severity: Medium

CVSS v3

Base Score: 6.5

Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Severity: Medium

CVSS v4

Base Score: 7.1

Vector: CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N

Severity: High

EPSS

EPSS: 0.00016