Dräger Infinity M300 patient worn monitors with software version VG2.3.1 and earlier contain a network-based denial of service vulnerability that allows network-adjacent attackers to repeatedly trigger device reboots by sending malicious requests over the Infinity Network. Attackers can exploit this vulnerability to force the device into a fail state requiring manual restart, causing loss of wireless connectivity and interruption of patient monitoring functionality.
https://www.vulncheck.com/advisories/dr-ger-infinity-m300-vg2-network-based-denial-of-service
Published: 2026-06-02
Updated: 2026-06-04
Base Score: 6.1
Vector: CVSS2#AV:A/AC:L/Au:N/C:N/I:N/A:C
Severity: Medium
Base Score: 6.5
Vector: CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
Severity: Medium
Base Score: 7.1
Vector: CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
Severity: High
EPSS: 0.00016