CVE-2018-8868

medium

Description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

References

https://ics-cert.us-cert.gov/advisories/ICSMA-18-179-01

https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-6-28-18.html

Details

Source: Mitre, NVD

Published: 2018-07-03

Updated: 2025-05-22

Risk Information

CVSS v2

Base Score: 6.9

Vector: CVSS2#AV:L/AC:M/Au:N/C:C/I:C/A:C

Severity: Medium

CVSS v3

Base Score: 6.4

Vector: CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H

Severity: Medium

EPSS

EPSS: 0.00094