CVE-2018-19010

medium

Description

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.

References

Details

Source: Mitre, NVD

Published: 2019-01-28

Updated: 2026-06-17

Risk Information

CVSS v2

Base Score: 3.3

Vector: CVSS2#AV:A/AC:L/Au:N/C:N/I:N/A:P

Severity: Low

CVSS v3

Base Score: 6.5

Vector: CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Severity: Medium

EPSS

EPSS: 0.00206