CVE-2018-14781

medium

Description

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

References

Details

Source: Mitre, NVD

Published: 2018-08-13

Updated: 2025-05-22

Risk Information

CVSS v2

Base Score: 2.9

Vector: CVSS2#AV:A/AC:M/Au:N/C:N/I:P/A:N

Severity: Low

CVSS v3

Base Score: 5.3

Vector: CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N

Severity: Medium

EPSS

EPSS: 0.00094