Description

An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.

References

Details

Source: Mitre, NVD

Published: 2018-02-15

Updated: 2026-06-17

Risk Information

CVSS v2

Base Score: 5

Vector: CVSS2#AV:N/AC:L/Au:N/C:N/I:N/A:P

Severity: Medium

CVSS v3

Base Score: 5.3

Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L

Severity: Medium

EPSS

EPSS: 0.00892