CVE-2017-12716

medium

Description

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

References

Details

Source: Mitre, NVD

Published: 2018-04-25

Updated: 2026-06-17

Risk Information

CVSS v2

Base Score: 3.3

Vector: CVSS2#AV:A/AC:L/Au:N/C:P/I:N/A:N

Severity: Low

CVSS v3

Base Score: 6.5

Vector: CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Severity: Medium

EPSS

EPSS: 0.00044