CVE-2015-1012

high

Description

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

References

https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01

Details

Source: Mitre, NVD

Published: 2019-03-25

Updated: 2019-10-09

Risk Information

CVSS v2

Base Score: 5

Vector: CVSS2#AV:N/AC:L/Au:N/C:P/I:N/A:N

Severity: Medium

CVSS v3

Base Score: 7.5

Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Severity: High

EPSS

EPSS: 0.00143