CVE-2014-5434

critical

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

References

https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01

Details

Source: Mitre, NVD

Published: 2019-03-26

Updated: 2019-10-09

Risk Information

CVSS v2

Base Score: 5

Vector: CVSS2#AV:N/AC:L/Au:N/C:P/I:N/A:N

Severity: Medium

CVSS v3

Base Score: 9.8

Vector: CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Severity: Critical

EPSS

EPSS: 0.00248